SOIL BOX for Soil Resistivity Measurement
Cat. #37008(Large)
Price $147.00
Cat. #37006(Small)
Price $137.00
For quick and accurate measurement of resistivity
of a sample of soil, water, or other electrolyte. For use in the
field or laboratory.
M. C. Miller soil boxes can be used to satisfy both the ASTM (G57) and AASHTO (T-288) Standard methodologies. The M. C. Miller soil boxes have a 1cm multiplier value if used in the 4 electrode configuration (ASTM Standard) and multiplier values of 0.67cm and 0.57cm for the small and large boxes, respectively, in the case of the two electrode configuration (AASHTO Standard).
Use the SOIL BOX with:
- Miller 400A or Miller 400D resistance meters
-
Any model MCM multi-combination meter.
-
Nilsson Model 400 soil resistance meter.
-
Megger.
-
Ammeter and Voltmeter or Milli-Ammeter and Milli-Voltmeter with external battery.
Made of Plexiglas:
|
LARGE
|
SMALL:
|
|
1.5"' wide by 8.75" long by 1.25" deep
|
1.16" wide x 4.39" long x .94" deep
|
|
3.937cm wide x 22.23cm long x 3.190cm deep
|
2.946cm wide x 11.15cm long x 2.3876 deep
|
- Corners: Rounded for easy
cleaning.
Soil Resistivity:
The dimensions of each box and the pin
spacings have been chosen so that the soil resistivity value (expressed in
Ohm-centimeter), is the same as the measured soil
resistance value (in Ohms). When used with any MCM
multi-combination meter (or separate volt and ammeter), and
external battery, the resistivity of the sample can be
determined as follows:
RESISTIVITY = [(Change in Potential Between Pins, or Delta V) / (Change in
Current, or Delta l)] x 1cm
Typical Meter Connections:
|
|
|
Resistivity of soil or water by box method using the MCM B3D digital multimeter..
|
Resistivity of soil or water by soil box method using an earth resistivity measuring
set, such as the Miller 400A Resistance Meter.
|
|
|
|
Resistivity of soil or water by soil box method using the MCM B3A2 analog
mutimeter.
|
Resistivity of soil or water by soil box method using a combination of meters
and batteries.
|
Soil Box Test Leads:
Set consisits of four(4) 2-foot long cables:
each with banana plug and pointed plug ends.
Notes:
-
When using multimeters, be sure readings on both meters are obtained
simultaneously.
-
Readings must be in volts and current must be in amps
(or mV and mA) to obtain result
in ohm-cm.
-
Multiplication factor for MCM Soil Boxes is 1cm.
-
When used with the Miller 400A Resistance meter or Megger, no
calculations are necessary because the multiplication factor
is 1cm. Therefore, the resistivity value of the sample
can be read directly off the meter dial. Refer to the
instruction sheet included with each meter for additional
information.